Assessment of IMU-derived lower-limb kinematics in transtibial amputees: A practical and accessible clinical workflow

IMPACT SIGNAL78/100
01

Information from the abstract

Clinical gait assessment in individuals with transtibial amputation is limited by the cost, technical demands, and space requirements of laboratory-based motion analysis. Although research-grade inertial measurement unit (IMU) systems provide a portable alternative, their hardware and software costs may restrict routine clinical use. Assessment in transtibial prosthesis users is further complicated by altered segment coordination, prosthetic component function, and limited anatomical landmarks. This study evaluated a transparent, independently implemented workflow using a consumer-grade IMU-based gait analysis system (cIMU) against a research-grade reference IMU system (rIMU). Thirty-two individuals with unilateral transtibial amputation completed five overground level-walking trials while wearing both systems, with sensors mounted in standardized adjacent positions. For the cIMU workflow, segment-orientation outputs were processed using custom Python scripts to derive bilateral hip, knee, and ankle kinematics, whereas rIMU kinematics were generated using the system's proprietary processing software. Inter-workflow agreement and repeated-trial consistency were assessed using Statistical Parametric Mapping (SPM), Bland-Altman analysis, intraclass correlation coefficients (ICCs), intra-subject variability, and paired comparisons across three anatomical planes. Repeated-trial consistency was high for hip kinematics across all three anatomical planes and for sagittal-plane knee and ankle kinematics, with average-measure ICCs ranging from 0.945 to 0.996. Inter-workflow agreement was strongest for sagittal-plane measures, with Bland-Altman mean differences within 2.5° for sagittal-plane knee and ankle measures. Wider limits of agreement were observed for several frontal- and transverse-plane measures and distal-joint variables. SPM identified significant waveform differences across multiple gait-cycle phases, particularly for frontal- and transverse-plane measures, while paired comparisons showed that differences in discrete measures were concentrated in non-sagittal and distal-joint variables. The cIMU workflow may be most useful for repeated-trial assessment of broad sagittal-plane gait patterns, whereas frontal- and transverse-plane kinematics and distal-joint measures should be interpreted cautiously. This accessible workflow may provide a practical option where research-grade gait-analysis systems are unavailable.

02

Why this record is monitored

This record has an Impact Signal of 78/100 based on recency, source, collaboration, and bibliographic signals. It prioritizes monitoring and is not a judgment of research quality.

Related topics: Prosthetics and Rehabilitation Robotics · Balance, Gait, and Falls Prevention · Muscle activation and electromyography studies

03

Thai researcher and institutional participation

Jutima Rattanakoch · Weerawat Limroongreungrat · Udomporn Manupibul · Harit Apidech · Wisavaporn Niamsang · Mintra Suphab · Thanaphat Laucha · Manunchaya Samala · Siriraj Hospital · Mahidol University · Chulabhorn Hospital

04

Data limitations

This page is a bibliographic record based on abstract-level information, not a full analysis or quality assessment. Verify the DOI and original article before citation.